VF counter

Share
Tracing N° 1
Biotronik Devices: ICD Field: Counter
Patient

A 43-year-old man implanted with a Lumax 340 VR-T single-chamber ICD for severe ischemic cardiomyopathy.

Graph and trace

Telemedicine tracing: 3 channels are available; the markers with the time intervals, the shock channel (FF: far field) between the ventricular lead coil and the pulse generator, the right ventricular sensing channel (V).

  1. sinus rhythm;
  2. polymorphic ventricular arrhythmia with extremely wide QRS complexes; short interval sensed in the VF zone;
  3. undersensing of a ventricular interval explaining that the ensuing interval is classified as VS;
  4. interval classified as VS in the absence of undersensing, the ventricular interval being slower than the programmed VF zone;
  5. the VF counter is full (programmed at 12/16); start of the capacitor charge;
  6. no therapy was delivered;
  7. end of the episode;

Programmer tracing: the 3 channels are the same as for the telecardiology tracing.

  1. start of the capacitor charge (black line);
  2. ventricular undersensing explaining the intervals classified as VS;
  3. double counting of the QRS complex;
  4. interruption of the capacitor charge following the sensing of 3 out of 4 intervals classified as VS; spontaneous interruption of the arrhythmia explaining the absence of delivered therapy.
Comments

This first tracing allows detailing the operation of the VF counter used in BiotronikTM ICDs. It is a probabilistic counter (fast X/Y intervals) with various programming options depending on the age of the device (for the Iperia 6/8, 8/12, 10/14, 12/16, 16/20, 18/24, 20/26, 22/30, 24/30, 30/40 ICD platforms). Ventricular fibrillation is, by definition, a rapid, disorganized, chaotic arrhythmia, with low and/or variable amplitude ventricular signals. All of these characteristics favor the risk of undersensing with sometimes amplitude signals below the sensing threshold (nominal value of 0.8 mV) and often a high beat-to-beat variability in amplitude which may detract the sensing circuit given the use by the ICD of a sensitivity level which adapts according to the amplitude of the preceding signal. The different counter values ​​were chosen so as to obtain an optimal balance between the accurate sensing of ventricular fibrillation (necessary tolerance of a certain number of long pseudo-intervals generated by undersensing) and the need not to fill the counters in the presence of T wave, P wave or R wave oversensing (frequently associated with a 50% short interval ratio).

In this patient, the VF counter was programmed at 18/24 (75% ratio) and the tracing showed 2 intervals classified as VS during the initial sensing without any significant impact on the sensing of the episode, the probabilistic counter tolerating a maximum ratio of 25% of long intervals.  The first VS interval was related to an intermittent undersensing, the second with a slower interval than the limit of the lower zone of the programmed VF zone. The detection quality of a ventricular arrhythmia is multiparametric and is contingent on:

  1. the filters and the amplification which vary according to the manufacturers;
  2. the amplitude of the ventricular signals with a limit set by the programmed sensitivity (nominal value 0.8 mV for BiotronikTM devices);
  3. the variability in signal amplitude: certain characteristics of the adjustment of the sensitivity level during the cardiac cycle (adjustment percentage and adjustment delay) are independently programmable;
  4. the number of intervals required to fill the counter which is programmable; as seen previously, the percentage of fast intervals (X/Y ratio) also varies slightly depending on the programming (66 to 80%);
  5. the limit of the programmed zones and the type of counter used; one of the features of the devices of this manufacturer is that the counting method differs completely between the VT zone and the VF zone.