Discrimination

Biotronik Device: ICD Field: Discrimination

Biotronik paths associated with this specificity

48-year-old man implanted with a Lumax 540 VR-T single-chamber ICD for ischemic cardiomyopathy with altered ejection fraction; event report (yellow color) in the setting of a classified VT1; programming of a VT1 zone (400-250 ms) with effective discrimination (onset 20%, stability 24 ms).

76-year-old man implanted with a Lumax 540 VR-T single-chamber ICD for dilated cardiomyopathy with ejection fraction of 25%; event report (yellow color) in the setting of a classified VT1; programming of a VT1 zone (350-290 ms) with effective discrimination (onset 20%, stability 24 ms).

Patient implanted with an Itrevia 5 VR single-chamber ICD; programming of 3 discrimination parameters (sudden onset 20%, stability 24 ms, MorphMatch ON); episode diagnosed as SVT in the device memory.

63-year-old man implanted with a Lumax 740 DR-T single-chamber ICD for ischemic cardiomyopathy; recording of a VT episode treated with an ATP burst; programming of a VT1 zone with SMART discrimination.

60-year-old man implanted with a Lumax 340 DR-T dual-chamber ICD for dilated cardiomyopathy with VT episodes; event report (yellow color) in the setting of a classified VT2.

 

61-year-old man implanted with a Lumax 740 DR-T dual-chamber ICD for ischemic cardiomyopathy with altered ejection fraction; event report (yellow color) in the setting of a classified SVT.

61-year-old man implanted with a Lumax 340 HF-T triple-chamber ICD for dilated cardiomyopathy with left bundle branch block; event report (yellow color) in the setting of a classified SVT.

59-year-old man implanted with a Lumax 740 DR-T dual-chamber ICD for dilated cardiomyopathy with VT episodes; event report (yellow color) in the setting of a classified VT1 and a SVT.